Download Inhoudsopgave Inhoud Print deze pagina

Gram BioCompact II RF210 Gebruiksaanwijzing pagina 111

Inhoudsopgave

Advertenties

Appendix
BioCompact II
Tolerances
REFRIGERATORS (RR)
210 (Solid door)
210 (Glass door)
310 (Solid door)
310 (Glass door)
+/- 3K
410 (Solid door)
410 (Glass door)
610 (Solid door)
610 (Glass door)
FREEZERS (RF)
210 (Solid door)
310 (Solid door)
**
410 (Solid door)
610 (Solid door)
*
The temperature span between the initial temperature and the end temperature
in the hold-over test P-13,14.
**
Please contact your local distributor for current information.
Note:
RR: Ambient temperature +25 °C
Setpoint temperature
RF: Ambient temperature +25 °C
Setpoint temperature -20 °C
Name:
Conducted by:
Inspected / verified by:
Door
Permissable
opening
deviation
recovery
time
(minutes)
4
35 %
4
10 %
4
35 %
4
20 %
8
15 %
7
10 %
6
35 %
6
35 %
15
20 %
11
10 %
19
10 %
10
20 %
+5 °C
Model: ________________________
Pull-
Permissable
down
deviation
(minutes)
23
5 %
28
5 %
24
5 %
28
5 %
21
5 %
22
5 %
15
5 %
**
5 %
53
10 %
62
5 %
71
10 %
40
10 %
Signature:
b i o l i n e
Hold-over
Hold-
Permissable
range*
over
66
43
55
42
5 °C → 10 °C
47
33
**
**
56
56
-20 °C → -10 ° C
42
**
Approved
Date:
(Yes / No):
SN: ________________________
deviation
5 %
5 %
5 %
10 %
10 %
5 %
10 %
10 %
5 %
5 %
10 %
10 %
111

Advertenties

Inhoudsopgave
loading

Deze handleiding is ook geschikt voor:

Biocompact ii rf310Biocompact ii rf410Biocompact ii rf610

Inhoudsopgave