Technical Information
Air Cooling Kit
Figure 24
294/306
Connect the FAN cable both on the FAN side and on the integrated
controller side (the cable must lay between the foreline flange and the
bottom side of the plastic Air cooling kit)
Fix the cable by locking the fixing clamps. Leave a ripple of about 5 – 6 cm
in length between the cable connector (FAN side) and the first locking
clamp (see Figures below)
Turbo-V 903 Pumping System User Manual / 87-901-016-01
Locking clamp
Locking clamp