Backwash procedure
5.1
Operation
• Begin backwashing by pressing the ON/OFF button on the
eBase Controller.
The processing time for 5 ml is at least 10 min.
5.2
End
After rinsing, the controller switches off automatically. If liquid is still present
in the funnel, please restart.
Note:
Aerosol may still escape from the easycare filling funnel at the end of
the rinsing process. This is normal. The eBase Controller can then also
be switched off manually by pressing the ON/OFF button.
6

Disassemble the easycare cleaning aid

• Disconnect the connection cord from the easycare reservoir and the
controller.
• Open the easycare reservoir and remove the aerosol head by pressing
on the two lateral engaging hooks (for disinfection of the aerosol head,
see Chapter "Cleaning and Disinfection" in the eFlow
Use).
• Discard the rinsing solution from the easycare reservoir.
7

Clean the easycare cleaning aid

After each use, clean with warm tap water and dishwashing soap. Then dry
thoroughly and store opened and turned over on a clean surface.