1. VOLTAGE /CURRENT Measurement
1-1 Voltage testing procedures
Step 1. Plug the Black test lead into the COM terminal and
the Red test lead into the V-Ω terminal.
Step 2. Set the rotary switch to the VA position.
Step 3. Connect the test leads to the device to be measured.
Step 4. Read the value from digital display.
- If the measured voltage is greater than 42 V DC or AC RMS,
the display will show the " " symbol, and the buzzer beeps
thrice.
- If the measured voltage exceeds 1000 V DC or AC RMS, the
display will show the "OL" symbol.
- If the measured frequency exceeds 1000 Hz , the display
will show the " out. F " symbol.