3. Insert the left connector into the new tube in its housing while making sure
that the rounded part is positioned inward.
4. Then turn the roller-holder clockwise so that the tube is inserted in its
housing.
5. Insert the right connector in its housing and tighten the two front screws to fix
the transparent cover.
Pump wintering:
Before shutting down the system for wintering or anyway for a long period, dose
clean water to rinse the tube then rotate the roller-holder clockwise to position it as
shown in the figure.
pH and ORP Electrodes
Typically, it is recommended to clean the electrodes when the response is slow or
measurements are not reliable and when they have been used for a long time,
especially in aggressive solutions, pollutants, very acidic, or very alkaline
environments. W'eau provides a kit of solutions for the cleaning and storage of pH
and ORP electrodes. The kit includes three solutions:
● Solution A: Dip the electrode in this solution for cleaning it.
● Solution B: Use this solution to rinse the electrode before and after cleaning.
● Solution C: Storage solution to be used for filling the electrode protective cap
when the electrode is not used (wintering closure of the plant).
Accessories and Spare Parts
Item
pH Electrode
ORP/Pt Electrode
ORP/Au Electrode
Description
Combined pH electrode with plastic body, 2.5 m
cable, and BNC
Combined ORP electrode with platinum sensor,
plastic body, 2.5 m cable, and BNC connector
Combined ORP electrode with gold sensor, plastic
body, 2.5 m cable, and BNC connector
Part No.
80094112
80194112
80194312