APPLICATION INFORMATION

EN
A P P L I C AT I O N I N F O R M AT I O N
1. Before ap p ly in g t he C r y o/ Cu f f,
p re pare the co o l er (F ig u re 1 )
Connect the blue tube to cooler.
Add cold water to line inside cooler.
Fill with ice.
Lay insulation disk on top of ice. Attach cooler lid snugly.
Allow five minutes with occasional shaking to chill water.
2. Alwa ys a pply E MPT Y C r y o /Cu ff
( Fi gure 2 )
Use only Aircast Cryo/Cuffs and ArcticFlow Cuffs with Aircast
Cryo/Cuff IC Unit. Cuff is the Applied Part.
Carefully and completely read instructions supplied with
the specific Cryo/Cuff prior to application. Please follow the
instructions below to continue cooler preparation.
3. Fill a n d pres s u r ize C ry o/ Cu f f
( Fi gure 3 )
To fill cuff: Connect blue tube to the Cuff. Raise cooler no more
than 15" (38cm) above the cuff for approximately 30 seconds
while Cryo/Cuff fills.
To correctly position cooler: Place cooler on stable surface.
The cooler base should be even with the Cuff.
NOTE: To AVOID excessive pressure during use, DO NOT raise
the cooler higher than 15" (38 cm) above the Cryo/Cuff.
To turn cooler on: Plug power supply into wall outlet and
connect cord to lid.
To turn off the equipment disconnect the power supply.
To refill cooler: When Ice has melted , unplug the cooler and
repeat items 1-5 under Step 1.
For specific cuffs, please refer to instructions for installation of
each cuff and follow procedure above.