Optima Turn/English
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of
this device should make sure it is used in such an environment.
Professional
Basic EMC
healthcare
standard
facility
environment
Electrostatic
±8 kV contact
Discharge(ESD)
± 2, 4, 8, 15 kV air
IEC61000-4-2
Electricalfast
±2 kV, 100 kHz for power supply line
transient/ burst
±1 kV, 100 kHz for input/output line
IEC61000-4-4
Surge
± 0.5, 1 kV (line to line) ±
IEC61000-4-5
0.5, 1, 2 kV (line to earth)
Voltagedips, short
Voltage Dips:
interruptionsand
i) 100% reduction for 0.5 period,
voltage variations
ii) 100% reduction for 1 period,
on powersupply
iii) 30% reduction for 25/30 period,
input lines
Voltage Interruptions:
IEC61000-4-11
100% reduction for 250/300 period
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz –80
Conducted RF
MHz
IEC 61000-4-6
6 Vrms
bands
between 0,15 MHz
Immunity Test Levels
HOME
HEALTHCARE
ENVIRONMENT
30 A/m
3 Vrms
0,15 MHz –80
MHz
inISM
6 Vrms in
and
radio
Compliance
Levels
±8kV
contact
± 2, 4, 8, 15
kV air
±2 kV, 100
kHzf
o
power
supply line
± 0.5, 1kV
(line to line)
230 Vac
30 A/m
3 Vrms
0,15 MHz–
80 MHz
ISM
6 Vrms in
amateur
ISMand
bands
amateur
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Mains power quality
r
should be that of
atypical commercial or
hospital environment
Mains power quality
should be that of
atypical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of this device
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Portable and mobile RF
communications
equipment should be
used no closer to any
part of this device,
including cables, than
16