Safety instructions & general use
Use of products
The following products are covered by this operating manual:
QS5, QS10, QS12, QS13, QS18, QS25 & QSx variants
The products listed above are a series of ultrasonic cleaning baths designed
for indoor use by a professional user.
When used to clean surgical/medical instruments, the products listed above
are classed as medical devices.
Any serious incident that has occurred in relation to the device should be reported to
the manufacturer and the competent authority of the Member State in which the user
and/or patient is established.
Electrical Safety
Connect to a 230VAC fully earthed supply via a 3pin plug.
It can be dangerous to operate an ultrasonic bath without an earth
connected.
The Q-Series machines are rated at up to 1050W maximum.
The detachable mains lead is rated at 250V 6A. NEVER FIT A LEAD WITH A
LOWER RATING.
The mains plug is fitted with a 5A fuse. NEVER FIT A FUSE OF A HIGHER
RATING.
The IEC electrical socket on the rear of the unit is fitted with 2x T5A (time-
lag) fuses. NEVER FIT A FUSE OF A HIGHER RATING.
The detachable mains lead is the disconnect device and should remain
accessible while the machine is in use.
Ensure that excess mains cable is stored neatly.
General use
Do not place fingers, hands, or other parts of the body in the liquid in the
bath when it is sonicating.
Fill the Q-Series with cold water. This will ensure that the machine runs its
automatic de-gas cycle (see page 27 for the importance of degassing).
Always ensure the liquid is above the level sensor when in operation.