1. Turn the rotary switch to the resistance measuring
range (CAP/Ω) and select resistance measurement
using the MODE button ('MΩ' appears on the display).
2. Insert the plug of the red measuring cable in the V/Ω
measuring socket and the black measuring cable in the
COM measuring socket.
3. Connect the measuring cables to the object to be mea-
sured. When measuring at the highest range, the
measuring instrument may under certain circumstan-
ces require a little time to display a stable value. This is
due to the principle of measurement and is not a mal-
function. In the case of very small resistance values
(400 Ω range), the internal resistances of the measu-
ring cables and sockets can lead to an incorrect display.
The resistance value displayed when the measuring tips
are connected together is noted and subtracted from
the subsequent measured values.
All manuals and user guides at all-guides.com
Diode test
This function enables semiconductor paths to be tested for
continuity in one direction and blocking in the other.
1. Turn
2. Insert the plug of the red measuring cable in the V/Ω
3. Connect the measuring cables to the object to be mea-
Operating Manual – Digital Multimeter BE 50
Before measuring resistance, continuity, diodes or
capacitance, switch off the current in the circuit
and discharge all capacitors.
the
rotary
switch
and select diode test using the MODE button ( is dis
played „
").
socket and the black measuring cable in the COM so-
cket.
sured, e.g. a diode. If over-range (OL) is displayed im-
mediately, swap over the measuring cable connections
on the object to be measured.
) /
to
„CAP/Ω/•
"
)
)