UK
Problem
The recording does
not start.
The recording stops
prematurely.
24 MU_BRIZZY_REV6 / 07-2019 ©
Possible cause
The mandibular movement sensor is
not properly connected to the device,
or is damaged.
The battery is empty.
The memory is full.
The USB cable is connected to the
device.
The battery has not been sufficiently
charged prior to the recording.
The battery is so worn out that even
if it is completely recharged before
each recording, it is not sufficient.
There is no more space in memory
for new data.
Solution
Check that the mandibular move-
ment sensor is properly connected
to the device. If damaged, check, if
possible, with another sensor.
Recharge the battery.
Connect the device to the computer
and import the data stored in
memory.
Disconnect the USB cable.
Recharge the battery completely as
soon as its level is low.
Battery life depends on usage and
different factors such as operating,
charging, and storage tempera-
tures. Follow the instructions in this
manual to optimize the battery life.
If the battery has worn out, contact
Nomics or your local distributor.
It is recommended to import the
recording stored in memory before
starting the next one. This will
empty the memory and free up
space for new data.
Nomics s.a. All rights reserved