6. MEASUREMENT PROCEDURE

Read the entire instructions before using the kit.
 Remove the cap from a vial of MI490A-0 Peroxide reagent.
 Use the graduated syringe to add exactly 1 mL of oil. For a correct use of the syringe,
please see GENERAL TIPS FOR AN ACCURATE MEASUREMENT.
 Add the sample to the vial and replace the cap.
 Mix by inverting the vial twice.
 Turn on the meter by pressing the ON/OFF key.
When the LCD displays "---", it is ready.
 Place the vial into the instrument.
 Press the ZERO key and "--- " will blink on the
display.
 After a few seconds the display will show "-0.0-".
The meter is zeroed and ready for measurement.
Remove the vial.
 Remove the cap from the vial and add one packet
of MI490B-0 Peroxide reagent.
 Replace the cap and press TIMER to start the
countdown.
 Mix vigorously for 1 minute.
 Insert the vial into the instrument and wait for
5 minutes.
Note: 30 seconds before countdown finishes,
invert the vial twice.
MEASUREMENT PROCEDURE 9