Electromagnetic compatibility (EMC, IEC 60601-1-2:2007 3rd Edition, Table 2)
Electromagnetic Immunity
The Medela Apgar Timer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Medela Apgar Timer should assure that they are used in such
environment.
Immunity Tests
IEC 60601
test level
Electrostatic
± 6 kV contact
Discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV
transient/burst
for power
IEC 61000-4-4
supply lines
± 1 kV
for input /
output lines
Surge
± 1 kV
IEC 61000-4-5
differential mode
± 2 kV
common mode
Voltage dips,
<5 % UT
short interrup-
(>95 % dip in UT)
tions and voltage
for 0.5 cycle
variations on
power supply
40 % UT
input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
All manuals and user guides at all-guides.com
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV
for power
supply lines
± 1 kV
for input /
output lines
not applicable
± 1 kV
differential mode
± 2 kV
common mode
not applicable
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
not applicable
3 A/m
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humi-
dity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user of the electric breast
pump Freestyle requires continued
operation during power mains in-
terruptions, it is recommended that
the electric breast pump Freestyle
is powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.