T T echnical description according to EN 60601-1-2
The Memmert CO
2
below. The customer or user of the CO
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Volage fluctuations / flicker
emissions IEC 61000-3-3
NOTE
The system impedance at the interface point according to IEC 61000-3-11 should not ecxeed
0.248+j0.155 Ohm resp. Zsys = 0.29 Ohm.
The Memmert CO
2
below. The customer or user of the CO
environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
tranient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
U
NOTE
is the mains voltage prior to application of the test level.
T
D39699 | Stand 01/2020
Guidance and manufacturer's declaration – electromagnetic emissions
Incubator ICOxxmed is intended for use in the electromagnetic environment specified
Incubator ICOxxmed should assure that it is used in such an
2
Compliance
Group 1
Class B
Class A
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
Incubator ICOxxmed is intended for use in the electromagnetic environment specified
Incubator ICOxxmed should assure that it is used in such an
2
IEC 60601
test level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input /
± 1 kV for input /
output lines
output lines
± 1 kV differential
± 1 kV differential
mode
mode
± 2 kV common
± 2 kV common
mode
mode
U
< 5 %
< 5 %
T
U
(>95 % dip in
)
(>95 % dip in
T
for 0,5 cycle
for 0,5 cycle
U
40 %
40 %
T
U
(60 % dip in
)
(60 % dip in
T
for 5 cycles
for 5 cycles
U
70 %
70 %
T
U
(30 % dip in
)
(30 % dip in
T
for 25 cycles
for 25 cycles
< 5 %
U
< 5 %
T
U
(>95 % dip in
)
(>95 % dip in
T
for 5 s
for 5 s
3 A/m
Not applicable
Electromagnetic environment - guidance
The CO
Incubator ICOxxmed uses RF energy only for ist
2
internal function. Therefore, ist RF emissions are very low
and are not likely to cause interference in nearby
electronic equipment.
The CO
Incubator ICOxxmed is suitable for use in all
2
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Compliance
Electromagnetic environment –
level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
enviroment.
Mains power quality should be that of a
typical commercial or hospital
enviroment.
U
Mains power quality should be that of a
T
U
)
typical commercial or hospital
T
enviroment.
U
If the user of the CO
T
U
)
ICOxxmed requires continued operation
T
during power mains interruptions, it is
recommended that the CO
U
ICOxxmed will be powered from an
T
U
)
uninterruptible power supply.
T
U
T
U
)
T
guidance
Incubator
2
Incubator
2
Bijlage