T T echnical description according to EN 60601-1-2
The CO
Incubator ICOxxmed is intended for use in electromagnetic enviroment in which radiated RF
2
disturbances are controlled. The customer or of the CO
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CO
according to the maximum power of the communications equipment.
Rated maximum output
power of trtansmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
d
distance
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, whre
transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and poeple.
Description of terminal
LAN port
Warning!
D39699 | Stand 01/2020
Recommended separation distances between portable and mobile
RF communications equipment and the Memmert steriliser type S..
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
= 1,2 √
0,12
0,38
1,2
3,8
12
P
is the maximum output power rating of the transmitter in watts (W) according to the
Listing of cables and maximum length of cables
The use of other cables may result in increased emissions or
decreased immunity of the Memmert CO
Incubator ICOxxmed can help prevent
2
Incubator ICOxxmed as recommended below,
2
m
80 MHz to 800 MHz
P
d
= 1,2 √
0,12
0,38
1,2
3,8
12
Type of cable
RJ45 CAT 6
800 Mhz to 2,5 GHz
P
d
= 2,3 √
0,23
0,73
2,3
7,3
23
Maximum length of cable
m
2
Incubator ICOxxmed
2
Bijlage
P