13. EMC / Technical description
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The electric breast pump Freestyle is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breast pump Freestyle should assure
that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
186
186
All manuals and user guides at all-guides.com
IEC 60601
Compliance
test level
level
3 Vrms
150 kHz to
80 MHz
3 Vrms
3 V/m
3 V/m
80 MHz to
2,5 GHz
Electromagnetic
environment – guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the electric
breast pump Freestyle, including ca-
bles, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
d = 1,2√P
d = 1,2√P
80 MHz to 800 MHz
d = 2,3√P
800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manu-
facturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey,
a
should be less
than the compliance level in each fre-
quency range.
b
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol: