To toggle between information displayed on the third LCD line, use
the  keys.
 If the reading is out of range, the full-scale value is displayed
blinking.
8.4. WARNINGS & MESSAGES
Messages Displayed During Calibration
 If the reading exceeds expected value, "WRONG STANDARD"
message is displayed and calibration can not be confirmed.
Check that correct calibration solution has been used and / or
clean the probe. See PROBE MAINTENANCE section for details.
 When using ATC mode, if the temperature of the solution
is outside the accepted interval, the "WRONG STANDARD
TEMPERATURE" message is displayed. The °C tag and the
temperature are displayed blinking.