10.TECHNICAL DATA
10.1 General technical data
Dimensions
• Depth
• Width
• Height
Weight
Temperature range
Height of figures
Power supply
• Batteries
- Supply voltage
- Battery type
• Optional power supply unit (model-
dependent)
- Supply voltage
- Insulated device, protection class II
(EN 60601-1)
Power consumption
• with wireless module disabled
• with wireless module activated
2. Hold down any key and start up the scales.
The current content of the calibration counter will
flash in the display for a few seconds.
3. Compare the output content of the calibration
counter with the figure on the calibration counter
sticker.
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, verifi-
cation is necessary. Contact your service partner or the
seca customer service department. Should verification
be necessary, the seca calibration counter sticker
shown above will be replaced by the adjacent verifica-
tion sticker to identify the calibration counter reading.
This sticker will be secured with an additional seal by a
person authorised to perform verification. The verifica-
tion counter sticker can be ordered from seca Service.
Technical data for
seca 704
+10 °C to +40 °C
6 x type AA, 1.5 V
approx. 32 mA
approx. 50 mA
Technical data • 65
English
520 mm
360 mm
930 mm
17 kg
25 mm
9 V
12 V