Store weight value (HOLD)
You can also continue displaying the weight determined once the load has been removed.
This makes it possible to look after a patient first, before having to note the weight.
– Press the HOLD/TARE key whilst the
load is on the scale.
"HOLD" appears in the digital display. The
value is stored until the scale switches off
automatically.
The triangle with the exclamation mark in-
dicates that this frozen display is an addi-
tional facility which is not subject to
calibration.
– The HOLD function can be switched off
by pressing the HOLD key again.
0. 0
appears in the display if the scale has
no load on it, otherwise the current weight.
The "HOLD" display vanishes.
Note:
The HOLD and TARE functions can also be used simultaneously.
Weighing with pre-Tare
If the pre-Tare function is activated, a stored weight value is deducted from the currently
measured weight. This function is useful, for example, if the weight of a wheelchair is not
to be displayed.
Three different tare weights (wheelchair weights) can be stored.
– Press the Start key.
se(a
8. 8 . 8 . 8 . 8 .
,
and
secutively in the display. After that, the
scale is automatically set to zero and
ready for operation.
– Press the FUNC key.
The last function used appears in the
display.
An arrow on the left-hand side of the
display indicates which memory is set.
– Use the arrow keys to select the de-
sired memory for storing the additional
weight or where a weight may already
be stored.
– Then press FUNC. In the display you
will see the tare value from the memory.
22
0. 0 0
. appear con-