Electromagnetic Compatibility
The device meets the requirements of EN 60601-1-2. All the accessories also meet the
requirements of EN 60601-1-2 when in use with this device.
Warnings and Cautionary Advices
Using accessories other than those specified in this manual may result in increased
⚫
electromagnetic emission or decreased electromagnetic immunity of the equipment.
The device or its components should not be used adjacent to or stacked with other
⚫
equipment.
The device needs special precautions regarding EMC and needs to be installed and put into
⚫
service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the requirements of
⚫
CISPR.
When the inputted signal is below the minimum amplitude provided in technical
⚫
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of this device.
⚫
Other devices that have RF transmitter or source may affect this device (e.g. cell phones,
⚫
PDAs, and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission tests
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations /
Flicker Emissions IEC
61000-3-3
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
Compliance
Group 1
Class B
Class A
Complies
IEC60601 test
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 1 kV line(s) to
Electromagnetic environment - guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 1 kV line(s) to
123
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
atleast 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.