Electrode potential tolerance ±300 mV
RF measurement range
Accuracy
Measurement overview
Standards
Verification of measurement accuracy: SpO2 accuracy has been verified in human testing by
comparing samples to blood reference values measured with a CO oximeter. Pulse oximeter
readings are statistically distributed and approximately two-thirds of the readings are
expected to be within a certain accuracy range compared to co-oximeter readings.
SpO2 range
SpO2 accuracy (arms)
PR range
PR accuracy
PI range
Measured value overview
Report
Curve report
Storage capacity
30 to 250 bpm
±2 bpm or ±2 %, whichever is greater.
Heart rate is calculated based on an average of 5 to 30
QRS complex each.
Heart rate, QRS duration, ST segment, rhythm analysis
(normal ECG rhythm, high heart rate, low heart rate,
high QRS. Irregular ECG rhythm, analysis not possible).
Complies with the standards from ISO 80601-2-61
70 % to 100 %
80-100 %: ±2 %, 70-79 %: ±3 %
30 to 250 bpm
±2 bpm or ±2 %, whichever is higher.
0,5-15
SpO2, overview (normal blood oxygen saturation, low
blood oxygen saturation, analysis not possible).
Complete curve evaluation
100 data sets
31
EN