9. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0108 (Hamburg Calibration Office). Rec-
alibration is essential if one or more security marks are damaged or the calibration counter
does not match the number on the applicable calibration counter mark (cf. page 20).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
10. Technical data
Dimensions scale
width:
height:
depth:
Dimensions weighing platform
width:
height:
depth:
Weight
Temperature range
Calibration
Medical product as per Directive
93/42/EWG
Colour
Maximum load
Minimum load
Graduations
Accuracy at first calibration
0 to 100 kg:
100 kg to 300 kg:
Power supply
Model 675
890 mm / 35 ins
+10
+0.4
65
mm / 2.6
965 mm / 38 ins
800 mm / 31.5 ins
+10
+0.4
46
mm / 1.8
965 mm / 38 ins
approx. 23.5 kg
+10 °C to +40 °C
approved for medical use, Class III
Class I
silver grey/black
300 kg / 660 lbs
4 kg / 8.8 lbs
200 g / 0.5 lbs
± 100 g
± 200 g
mains unit
ins
ins
GB
25