directly onto the device.
 When cleaning the device with water, the temperature should be lower than
60°C.
 As to the fingers which are too thin or too cold, it would probably affect the
normal measure of the patients' SpO2 and pulse rate, please clip the thick
finger such as thumb and middle finger deeply enough into the probe.
 Do not use the device on infant or neonatal patients.
 The product is suitable for children above four years old and adults (Weight
should be between 15 kg to 110 kg).
 The device may not work for all patients. If you are unable to achieve stable
readings, discontinue use.
 The update period of data is less than 5 seconds, which is changeable
according to different individual pulse rate.
 Please read the measured value when the waveform on screen is equably and
steady-going. This measured value is optimal value. And the waveform at the
moment is the standard one.
 If some abnormal conditions appear on the screen during test process, pull out
the finger and reinsert to restore normal use.
 The device has normal useful life for three years since the first electrified use.
 The hanging rope attached to the device is made from Non- allergy material, if
particular group are sensitive to the hanging rope, stop using it. In addition,
pay attention to the use of the hanging rope, do not wear it around the neck
on the purpose of avoiding harm to the patient.
 The instrument does not have low-voltage alarm function, it only shows the
low-voltage. Please change the battery when the battery energy is used out.
 The instrument dose not have alarm function. Do not use the device in
situations where alarms are required.
 Batteries must be removed if the device is going to be stored for more than
one month, or else batteries may leak.
 A flexible circuit connects the two parts of the device. Do not twist or pull on
the connection.
1.4 Indication for Use
The Fingertip Pulse Oximeter is a non-invasive device intended for the spot-check
of oxygen saturation of arterial hemoglobin (SpO
pediatric patients in home and hospital environments (including clinical use in
internist/surgery, anesthesia, intensive care ect.). This device is not intended for
continuous monitoring.
2. Overview
The pulse oxygen saturation is the percentage of HbO
blood, so-called the O
concentration in the blood. It is an important bio-
2
) and the pulse rate of adult and
2
in the total Hb in the
2
19