Pulox PO-200 Gebruiksaanwijzing pagina 20

Pulsoxymeter
parameter for the respiration. For the purpose of measuring the SpO
and accurately, our company developed the Pulse Oximeter. At the same time,
the device can measure the pulse rate simultaneously.
The Pulse Oximeter features in small volume, low power consumption,
convenient operation and being portable. It is only necessary for patient to put
one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display
screen will directly show measured value of Hemoglobin Saturation.
2.1 Classification
Class II b, (MDD93/42/EEC IX Rule 10)
2.2 Features
A. Operation of the product is simple and convenient.
B. The product is small in volume, light in weight (total weight is about 50 g
including batteries) and convenient in carrying.
C. Power consumption of the product is low and the two originally equipped AAA
batteries can be operated continuously for 20 hours.
D. The product will enter standby mode when no signal is in the product within 5
seconds.
E. Display direction can be changed automatically, easy to view.
2.3 Major Applications and Scope of Application
The Pulse Oximeter can be used in measuring the pulse oxygen saturation and
pulse rate through finger. The product is suitable for being used in family,
hospital, oxygen bar, community healthcare, physical care in sports (It can be
used before or after doing sports and it is not recommended to use the device
during the process of having sport) and etc.
The product is not suitable for use in continuous supervision for patients.
The problem of overrating would emerge when the patient is suffering
from toxicosis which caused by carbon monoxide, the device is not
recommended to be used under this circumstance.
2.4 Environment Requirements
Storage Environment
a. Temperature : -40°C ~+60°C
b. Relative humidity : ≤95%
c. Atmospheric pressure : 500hPa~1060hPa
Operating Environment
a. Temperature : 10°C ~40°C
b. Relative Humidity : ≤75%
c. Atmospheric pressure : 700hPa~1060hPa
3. Principle
3.1 Principle of Measurement
Principle of the Oximeter is as follows: An experience formula of data process is
established taking use of Lambert Beer Law according to Spectrum Absorption