8. Cleaning, Transportation and Storage
 Please change the batteries when the low-voltage displayed on the screen.
 Please clean the surface of the device before using. Wipe the device with
medical alcohol first, and then let it dry in air or clean it by dry clean fabric.
 Using the medical alcohol to disinfect the product after use, prevent from
cross infection for next time use.
 Please take out the batteries if the oximeter is not in use for a long time.
 The best storage environment of the device is -40ºC to 60ºC ambient
temperature and not higher than 95% relative humidity.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry environment.
Humidity may reduce the useful life of the device, or even damage it.
9. Troubleshooting
Trouble
Possible Reason
The SpO2 and
1. The finger is not properly
Pulse Rate can
positioned
not be
2. The patient's SpO2 is too low to
displayed
be detected.
normally
The SpO2 and
1. The finger is not placed inside
Pulse Rate are
deep enough.
not displayed
2. The finger is shaking or the patient
stably
is moving.
1. The batteries are drained or
The device can
almost drained.
not be turned
2. The batteries are not inserted
on
properly.
3. The malfunction of the device
1. The device will power off
automatically when it gets no signal
The display is
within 5 seconds.
off suddenly
2. The battery is almost drained
away.
10. Key of Symbols
Symbol
Type BF applied part
Description
Solution
1. Place the finger properly and
try again.
2. Try again; Go to a hospital
for a diagnosis if you are sure
the device works all right.
1. Place the finger properly and
try again.
2. Let the patient keep calm
1.Change batteries
2. Reinstall batteries.
3. Please contact the local
service center.
1. Normal.
2.Change batteries
Symbol
WEEE (2002/96/EC)
24
EN
Description