Battery low
Approximately 30 minutes before the battery is fully empty an acoustic signal sounds and the
battery symbol [
] starts to blink� The acoustic signal can be muted but the battery symbol
continues to blink until the battery is empty� The set pressure is maintained but the battery should
be recharged as soon as possible�
Battery empty
The battery symbol will blink with an acoustical sound for 10 minutes prior to the battery fully
discharging, unless the pump is switched off earlier� The acoustic sound cannot be muted during
the 10 minutes� If Thopaz + is not connected to a power source the pump will switch off automati-
cally after 10 minutes� The negative pressure is not maintained after the pump turns off�
*
1� Battery empty
*
2� Battery almost empty
3� Battery is being charged (moving bars)
4� Battery is fully charged and connected to the mains
* Symbol blinks
12 |
Power supply and battery operation
Press simultaneously to acknowledge an acoustic alarm and
follow instructions on the display (see chapter 12 – Trouble-
shooting and alarms)�