Drying
After manual cleaning and disinfection, store the product on a clean and dry cloth and let it dry
or follow instructions for use of the disinfectant manufacturer�
Inspection
After reprocessing according to the steps described above, the device and its components
must be inspected for signs of degradation that may limit the useful life and/or performance
of the device, such as the following: visible corrosion, mechanical wear, abrasion, damage,
or deformation� Discard the device if any signs of degradation are evident�
Storage
Store the product dry and dust-free�
Disposables
This symbol indicates a single use product� This product is not intended to be reprocessed�
Reprocessing could cause loss of mechanical, chemical and/or biological characteristics�
Reuse could cause cross contamination�
46 |
General cleaning guidelines