Switch Thopaz + on
7�4
Important: Do switch Thopaz + on before connecting
the system to the patient.
1�
If the selftest is not successful, follow the troubleshooting
instructions on the display and refer to chapter 12 – Trouble-
shooting and alarms�
2�
Thopaz + will issue a new patient number�
yes
The therapy time and history data will start from
0 (recommended for a new patient)�
Note: Once the patient number reaches 100 it
automatically defaults back to 1.
no
The patient number remains unchanged (recom-
mended for the continued treatment of the same
patient)� The data history and the therapy time are
retained�
3�
1�1 Press [
] to switch
Thopaz + on� A selftest
starts�
1�2 Check acoustic signal
(beep)�
2�1 Confirm whether a new
patient is connected or
not� The patient number is
important for therapy time,
history data and data
transfer to the PC�
3�1 Check canister matching:
detected canister size vs�
attached canister�
No match: dispose
canister�
Preparation for use/operating instructions
| 27