 In outdoor measuring systems, the result of measurement may
be influenced by the ambient temperature.
 Insulate measuring elements outside the pipe adequately to
prevent ambient temperature influences.
 Fill the thermowell(s) with a heat-conductive fluid or paste to
achieve optimum thermal conduction.
2.4 Pressure test points
A straight male coupling which complies with DIN 2353 is provided on the
meter housing to act as a pressure test point, for example to connect a
pressure sensor. It is marked pm/pr and is designed for connecting Ø 6 mm
steel tubes to DIN EN 10305-1 (e.g. steel grade E235) or flexible pressure
tubes from Honeywell.
CAUTION!
 Danger to functional safety!
 Functional safety and reliability are ensured only if the material
combination of the union component and the pipe are
intermatched.
 The straight male coupling must not be connected to pipes
made of stainless steel or pipes made of non-ferrous materials.
NOTE!
 Only use original Parker-Ermeto or Voss pipe unions.
 We recommend that you contact our local Honeywell Customer
Service Centre for conversion work and when installing
additional devices.
2.5 Pulse generators/Encoders
The meter may be fitted with the following pulse generators or encoders:
Device type:
Low-frequency pulse generator (LF)
Low-frequency pulse generator (LF)
Low-frequency pulse generator (LF)
High-frequency pulse generator (HF)
Encoder
Table 2 | Overview of pulse generators
Further information on the pulse generators and encoders is provided in
the separate instruction manuals.
18
Manufacturer:
Elster GmbH
Elster GmbH
Elster GmbH
Pepperl & Fuchs
Elster GmbH
73021678h
© Elster GmbH | All rights reserved. Subject to modification
Device designation:
IN-Sxx
IN-Cxx
S1xRx
SJ2-N
ENCODER S1