P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
W W A A R R N N I I N N G G - - Do not modify or change this device. Service should only be completed by qualified personnel. Failure could
result in injury and void your warranty.
Remove product from service before you perform preventive maintenance. At a minimum, check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel
N N o o t t e e - - Consult your local regulations to dispose of electronic equipment.
Inspect the following items:
All fasteners are secure
Pump housing or components (hoses, power cords, or case) for cracks, holes, or damage
Bed frame pump hooks are not damaged
No air leaks from the pump or the attached connectors or hoses
Graphical user interface is not cracked or damaged
HEPA filter (replace each year)
Fuse
All functions on graphical user interface operate
Run diagnostic test (All okay)
Product serial number:
Completed by:
Date:
2874-009-005 Rev AB.2
21
EN