Ultrasound power
The left rotary knob is used to adjust the ultrasound
power from 10 % to 100 % (in 10 % increments).
Duty cycle
The duty cycle is the ratio between ultrasound
ON and ultrasound OFF over an interval of time
of 80 milliseconds. Set the duty cycle using the
arrow keys.
The last setting of the duty cycle is kept in memory
at power off.
The handpiece and probe error detection
is possible only with a duty cycle set at 100 %
(see Troubleshooting chapter).
Activation
Press the ultrasound foot pedal halfway to
activate the suction (pinch valve opens).
Press the foot pedal completely to activate
ultrasound handpiece, probe, and suction.
20
Snaga ultrazvuka
Lijevi okretni gumb upotrebljava se za namještanje
snage ultrazvuka od 10% do 100% (u koracima
po 10%).
Radni ciklus
se tipkama sa strelicama.
Posljednje namješteni radni ciklus sprema se u
100% (vidi poglavlje Uklanjanje kvarova).
Aktiviranje
Pritisnite pedalu za ultrazvuk do pola kako
biste aktivirali usisavanje (otvara se dozirni
ventil).
Pritisnite pedalu u potpunosti kako biste
usisavanje.
Ultrasound vermogen
De linker draaiknop wordt gebruikt om het
ultrasound vermogen in te stellen van 10 % tot
100 % (in 10 % stappen).
Taakcyclus
De taakcyclus is de verhouding tussen ultrasound
AAN en ultrasound UIT binnen een tijdsinterval
van 80 milliseconden. Stel de taakcyclus in met
de pijltjestoetsen.
De laatste taakcyclusinstelling wordt in het
geheugen opgeslagen als de unit wordt uitgezet.
Foutdetectie van het handstuk en de sonde
is uitsluitend mogelijk met een taakcyclusinstelling
van 100% (zie hoofdstuk Problemen oplossen).
Activeren
Druk het ultrasound bedieningspedaal half
in om de afzuiging te activeren (de knijpklep
gaat open).
Druk het bedieningspedaal volledig in om
het ultrasound handstuk, de sonde en de
afzuiging te activeren.
loading